IMPORTANT: This is a brief supplementary guide and does not replace the product’s current Instructions for Use (IFU). Always read and follow the IFU, as well as local workplace instructions and the individual’s care plan.
IMPORTANT: ZiboCare cannot assess the individual user’s clinical situation. Use of the product and assessment of the user’s needs must be carried out by qualified professional personnel in accordance with the applicable Instructions for Use (IFU), local guidelines, and the individual care plan.
Always use the latest IFU for the specific product and model number. In the event of a discrepancy between this guide and the IFU, the IFU takes precedence.
We have gathered information here on what to be aware of if you have a mattress with a pump and a power outage occurs.
In the event of a power failure, it is advisable to take the following 6 measures:
Check the patient’s position, comfort, and risk of pressure injuries, as well as the mattress’s current firmness.
Check the wall outlet, plug, and power cord. Do not disconnect the power cable from the pump.
As a general rule, hoses and connectors should remain connected.
The pump’s power failure alarm activates during a power outage. Handle and—if necessary—deactivate the alarm as described in the applicable Instructions for Use (IFU) or Quick Guide.
The alternating pressure function is inactive without power. Follow the patient’s care plan and local clinical guidelines regarding repositioning and skin assessment. Consider transferring the patient to a pressure-relieving foam mattress.
Regularly check whether the mattress is losing firmness and if the patient remains adequately supported.
IMPORTANT: The responsible healthcare professional must continuously assess the individual’s need for support, pressure relief, and potential relocation. Relocate the individual to a suitable alternative before the support becomes inadequate.
In the event of a prolonged power outage, a full-size mattress may gradually lose air, potentially resulting in the user receiving inadequate support from the mattress.
ZiboCare’s general guidelines state that a new full-size mattress, with a user weighing approximately 100 kg in a flat position, can retain air for at least 10 hours*.
The topper mattress must be used in conjunction with the associated safety mattress. Check that the safety mattress is correctly positioned and in good condition.
ZiboCare’s general guidelines state that a new overlay mattress, with a user weighing approximately 100 kg in a flat position, can retain air for at least 8 hours*.
Age, the individual’s weight, and damage to or degradation of the mattress can reduce its lifespan. Therefore, always assess the specific mattress and the individual.
Relocate the individual or implement an alternative solution if the overall support provided is no longer sufficient.
* This information is an indicative figure based on experience; it is not a guarantee and must not serve as the sole basis for clinical decisions regarding how long the individual can remain on the mattress without power.
As a general rule, this product group poses no risk, as the mattress can function both with and without a pump. If the pump is connected and a power outage occurs, the mattress will function as a pressure-relieving mattress without the risk of the user bottoming out.
You can get more general information about power outages from your local electricity company, or you can read more in the Emergency Management Agency's brochure “Prepared for Crises”.